Cleaning the ion guide assembly
4-59
3. Use the hook to carefully suspend the ion guide assembly in the glass
vessel so that the bottom of the assembly does not touch the bottom of
the vessel.
4. Add 1:1 methanol/water to the glass vessel until the ion guide assembly
is immersed completely.
5. Place the vessel in the ultrasonic bath for 30 minutes.
6. Carefully remove the ion guide assembly from its vessel, and blow-dry it
using inert, oil-free gas.
7. Inspect the ion guide assembly for persisting contamination. If
contamination is present, do as follows:
a. Use the wash bottle containing methanol to rinse the ion guide
assembly over the large beaker.
b. Blow-dry the ion guide assembly with inert, oil-free gas.
Caution:
To avoid vibration damage to the ion guide assembly,
ensure that the bottom of the assembly is not in contact with the
bottom of the glass vessel.
Caution:
To avoid recontaminating the ion guide assembly, wear
clean, chemical-resistant, powder-free gloves for the rest of this
procedure.
TP03412
Hook
Rear circuit board
carrier
Содержание Xevo TQD
Страница 12: ...xii...
Страница 20: ...xx Table of Contents...
Страница 23: ...Uses and compatibility 1 3 Xevo TQD TP03406...
Страница 24: ...1 4 Specifications and Operating Modes Xevo TQD with visor up TP03407...
Страница 42: ...1 22 Specifications and Operating Modes...
Страница 52: ...2 10 Preparing for Operation...
Страница 76: ...3 24 Changing the Mode of Operation...
Страница 194: ...4 118 Maintenance Procedures...
Страница 208: ...A 14 Safety Advisories...
Страница 234: ...B 26 External Connections...
Страница 238: ...C 4 Materials of Construction and Compliant Solvents...
Страница 250: ...Index 8...