3-18
Changing the Mode of Operation
4. On the NanoFlow source, loosen the stage retaining screw, pull the stop
screw, and slide the stage fully out of the enclosure.
5. Using both hands, fit the NanoFlow source enclosure to the two
supporting studs on the source adaptor housing.
6. Close the source enclosure door.
7. Connect a 1/16-inch PTFE tube between the mass-flow controller output
(mounted beneath the stage on the front of the NanoFlow source) and
your sprayer.
Tip:
For details regarding how to fit each sprayer, see the corresponding
reference:
•
Waters Universal NanoFlow Sprayer Installation and Maintenance
Guide
(part number 71500110107)
•
“Fitting a borosilicate glass capillary (nanovial)” on
•
Capillary Electrophoresis/Capillary Electrochromatography
Sprayer User's Guide
(part number 6666522)
8. Connect the probe cable to the high-voltage connector.
Caution:
To prevent the sprayer from colliding with the cone and
breaking, always retract the stage before installing the source
enclosure or closing the door.
Stop screw
Retaining screw
Содержание Xevo TQD
Страница 12: ...xii...
Страница 20: ...xx Table of Contents...
Страница 23: ...Uses and compatibility 1 3 Xevo TQD TP03406...
Страница 24: ...1 4 Specifications and Operating Modes Xevo TQD with visor up TP03407...
Страница 42: ...1 22 Specifications and Operating Modes...
Страница 52: ...2 10 Preparing for Operation...
Страница 76: ...3 24 Changing the Mode of Operation...
Страница 194: ...4 118 Maintenance Procedures...
Страница 208: ...A 14 Safety Advisories...
Страница 234: ...B 26 External Connections...
Страница 238: ...C 4 Materials of Construction and Compliant Solvents...
Страница 250: ...Index 8...