NanoFlow source
3-19
9. Connect the high-voltage cable to the instrument’s HV connection.
10. Slide closed the instrument’s source interface door.
Fitting a borosilicate glass capillary (nanovial)
Required materials
•
Chemical-resistant, powder-free gloves
•
Needle-nose pliers
•
Borosilicate glass capillary
•
Fused silica syringe needle or GELoader
®
tip
•
Fused silica cutter
Caution:
The NanoFlow stage contains a high-voltage interlock, so
the capillary voltage (the voltage applied to the sprayer assembly)
and the sampling cone voltage remain disabled until the sprayer is
pushed fully forward in the source.