4-56
Maintenance Procedures
3. Use the 3-mm Allen wrench to unscrew and remove the 4 screws
securing the PEEK ion block support to the adaptor housing, and
remove the ion block support.
4. If any of the O-rings show signs of deterioration or damage, dispose of
them in accordance with local environmental regulations.
Warning:
The O-rings can be contaminated with
biohazardous and/or toxic materials. Dispose of them
according to local environmental regulations.
Securing screws
Adaptor housing
PEEK ion block
support
Ion guide
assembly
Содержание Xevo TQD
Страница 12: ...xii...
Страница 20: ...xx Table of Contents...
Страница 23: ...Uses and compatibility 1 3 Xevo TQD TP03406...
Страница 24: ...1 4 Specifications and Operating Modes Xevo TQD with visor up TP03407...
Страница 42: ...1 22 Specifications and Operating Modes...
Страница 52: ...2 10 Preparing for Operation...
Страница 76: ...3 24 Changing the Mode of Operation...
Страница 194: ...4 118 Maintenance Procedures...
Страница 208: ...A 14 Safety Advisories...
Страница 234: ...B 26 External Connections...
Страница 238: ...C 4 Materials of Construction and Compliant Solvents...
Страница 250: ...Index 8...