Parameter
Screen
Description
Type
Filt (DC mode) DC SETUP, DC STAT
Filter settings: RAW (100 Hz), Off (10 Hz),
and 1 Hz to 0.001 Hz cutoff frequency, in
1, 2, 5 steps.
C
Filt (PULSE
mode)
PULSE SETUP, PULSE
STAT
Filter settings: Off and 0.5 Hz to 0.001 Hz
cutoff frequency, in 1, 2, 5 steps. (Fcut-off /
filter coefficients based on 1 Hz input
frequency in pulse mode).
C
Filt
RUN
Reflects the actual filter setting.
S
Firmware
SYSTEM
Displays firmware version.
S
Hold resume
RUN, SCAN STAT
Toggle—holds or resumes execution of
scan.
S
HOLD=0,1
RUN, SCAN STAT
Holds or continues execution of scan.
Toggles between 1 and 0. Pressing hold
again continues scan where it was held.
S
Ic
STAT (dc, pulse, scan),
RUN, NOISE
Displays the true, non-compensated cell
current, unaffected by auto zero or offset.
F
ID1 master
CONFIG
Sets sensor board 1 as master. When this
setting is set to ‘yes’ all parameter settings
from sensor board 1 are automatically
copied/transferred to all other sensor
boards present.
F
IO
CONFIG
Enter IO menu.
S
INJ=I/L
DC STAT, PULSE STAT
Displays or switches the position of the
injection valve, toggles between inject (I)
and load (L). If a manual injector with
position sensor is applied, it echoes the
position of the injector. If an electrically
actuated injector is used (optional), it is
possible to switch the injector with this
function button.
C
KEYB
TEST
Enters “KEYB” screen, for keyboard test.
Press 2x F1 to leave.
F
MARK
DC STAT, PULSE STAT
Triggers a marker signal on output.
F
MaxComp
DC SETUP, PULSE
SETUP1
Maximum cell current that can be
compensated for using auto zero.
S
Next
SEVERAL SCREENS
Enter next screen.
F
NOISE
TEST
Enters NOISE screen for performance
test.
F
December 16, 2021, 715007395 Ver. 00
Page 48