Figure 12–7: Noise-Stop Flow
Note:
If no significant drop in noise or cell current is observed, service or replace the cell.
If you still cannot solve the problem, contact your local representative.
12.4.1 Possible solutions for analytical problems
Bear in mind that analytical problems may also be caused by external influences like temperature
or unstable samples. Ensure that the application was running trouble-free before and that no
changes were made to the system. A number of causes and possible solutions for analytical
problems are listed here. Contact your local representative if you need further help.
Table 12–5: No detector response
Possible cause
Remedy
No power
Check line voltage setting, plug in power cord
Power switch off
Turn this switch ON (at the rear panel)
Faulty fuse
Replace fuse
Divergent mains voltage
Check line voltage
Cell disconnected, or switched off
Check connection
Output disconnected
Check connection
Fouled WE
Clean WE