Parameter
Screen
Description
Type
compensation will be applied, which may
be higher than the 0 Volt level.
Boot
SYSTEM
Displays boot firmware version.
F
CELL=ON/
OFF
DC STAT, PULSE STAT,
SCAN SETUP, SCAN STAT
Toggles between cell ON and OFF.
Confirmation is required—"switch cell on
(off)?” Switching on resets the clock to
0.00. Pulse mode: pulsation occurs as
long as the cell is on, irrespective of the
screen selected. Scan mode: potential E1
is applied.
F
Checksum
SYSTEM
Displays checksum
S
Comp
DC STAT, PULSE STAT
Toggles between ON and OFF, releases
auto zero offset. Switches ON if AZERO is
pressed. Affects auto zero compensation
only, not the % offset.
C
CONFIG
MAIN
Enters config screen.
F
Contrast
CONFIG
Sets the contrast of display.
C
Cyc
SCAN SETUP
Controls the nature of the cycle: half, full,
and continuous. “Half” means that the cell
potential runs from E1 to E2 and stops at
E2 (/| ). “Full” means that the cell potential
runs from E1 to E2, back to E1, and then
stops (/\). “Cont” means that the cell
potential runs from E1 to E2 and back to
E1 continuously (/\/\/\/\……..). Pressing
“STOP” or finishing the cycle sets the
potential to E1.
C
DIAG
MAIN
Enters Diag screen.
F
DISPL
TEST
Enters DISP screen for display test.
F
E1, E2, E3, E4,
E5
PULSE SETUP2, PULSE
SETUP3
Controls the cell potential settings of the
pulse. Can be set b2.50 and –
2.50 V in 10-mV steps. Can only be set or
changed after confirmation with the
Enter
button.
C
Ec
DC SETUP
Controls the cell potential; can be set
b2.50 and –2.50 V in 10-mV
steps. Can only be set or changed after
confirmation with the
Enter
button.
C
Ec
RUN (DC only), SCAN
STAT (during scanning)
Reflects the set cell potential. Displays the
actual cell potential in the scan mode.
S
December 16, 2021, 715007395 Ver. 00
Page 47