Data Rate
1 - 100 Hz in 1, 2, 5 increments, dependent on
filter setting
Noise
< 2 pA with dummy cell (load of 300 MΩ/470
pF) in 1 nA range, filter off, Ec +800 mV and
temperature of 35 °C.
10.4 3465 PULSE mode
Range
10 nA - 200 µA in 1, 2, 5 increments
Filter (ADF)
0.5 - 0.001 Hz in 1, 2, 5 increments
OFF: for unprocessed data
Potential (Ec)
-2.50 V to + 2.50 V in 10-mV increments
Data rate
1/(pulse duration) Hz
Waveform
Max 5 potential steps
Pulse times (t1-t5)
t1: 100 ms - 2000 ms; t2, t3, t4, t5: 0 - 2000 ms
in 10-ms increments
Sampling times (ts)
20 ms - [t1 - 60] ms
10.5 3465 PULSE mode 2
Range
10 nA - 200 µA in 1, 2, 5 increments
Filter (ADF)
0.5 - 0.01 Hz in 1, 2, 5 increments
OFF: for unprocessed data
Potential (Ec)
-4.90 V to + 4.90 V in 10-mV increments
Data rate
1/(pulse duration) Hz
Waveform
Free programmable multi-step waveform with
up to 30 time-potential (t, E) coordinates and
max. pulse duration of 4 s. Time points in 10-
ms increments.
Sampling time
Sampling interval is freely programmable,
defined by Begin and End markers.
10.6 3465 SCAN mode
Range
10 nA - 200 µA in 1, 2, 5 increments
December 16, 2021, 715007395 Ver. 00
Page 83