Warning:
Use only manufacturer-supplied cable to connect devices. Thoroughly connect
shielding to common ground. The manufacturer accepts no liability for damage, direct or
indirect, caused by connecting this instrument to devices with cables that do not meet
relevant safety standards.
2.2.4 Chemicals
Mobile phase and flush/storage solutions must be clean because they are in direct contact with
the working electrode of the electrochemical cell. High-purity chemicals, including water, are
prerequisites. All chemicals should be electrochemically clean, HPLC-grade or better. For water
used in the preparation of mobile phases, a purification apparatus is recommended that is able to
supply high-purity deionized water with resistivity of >18 Mohm*cm and low (<10 ppb) TOC level.
2.3 Unpacking
Inspect the transport box for possible damage when it arrives. Immediately inform the transport
company of any damage, otherwise it may not accept responsibility. Keep the transport box; it is
designed for optimum protection during transport and may be needed again. Carefully unpack the
system and inspect it for completeness and possible damage. Contact your supplier about
damage or if not all marked items on the checklist are included. Prior to shipment, your detector
was thoroughly inspected and tested to meet the highest possible demands. The results of all
tests are included.
See the checklist below for reference:
1. Delivery is in accordance with the order.
2. Delivery is undamaged.
3. All items on Start Up Kit list are included.
4. Certificates of performance are included:
a. Detector
b. Flow cells
1
5. See startup kit list for weblink to user manual.
To unpack the detector, lift it from its box by both hands. Never lift the detector by its front door;
lift it by the sides.
1 1
Flow cells are not part of the 3465 Detector and must be ordered separately.
December 16, 2021, 715007395 Ver. 00