• Stable, clean, flat, and smooth surface.
• Enough mechanical strength to hold at least the weight of the detector. A 3465 Detector
without cell weighs 14.4 kg (31.7 lbs). A fully dressed detector with flow cells, peripherals,
columns, and valves may weigh 20 kg or more.
• A 3465 Detector has the following dimensions: Depth 44.0 cm (17.3 inches), Width 22.0 cm
(8.7 inches), and Height 43.0 cm (16.9 inches). Take into account that additional space is
necessary on all sides to prevent obstruction of ventilation holes and allow sufficient heat
dissipation. Keep at least 15 cm free at the back. Keep at least 5 cm distance if there is
another device on one side. Keep at least 10 cm distance if there are devices on both sides.
2.2.3 Electrical and power requirements
The customer is responsible for providing appropriate electrical power and power outlets in the
laboratory.
1. The installation of electrical supplies and fixtures in the laboratory must comply with all
local regulations and safety standards.
2. The power source should exhibit minimal power transients and fluctuations. The AC main
supply voltage source should not fluctuate more than +/- 10% from nominal voltage. If your
main voltage is unstable (>10% of nominal voltage), use an Uninterruptible Power Supply
(UPS). The main supply must include a correctly installed protective earth conductor.
Notice:
To protect against power transients (voltage spikes and power surges),
it is recommended that you connect the equipment over an electrical surge
protector.
3. The 3465 Detector is equipped with a universal AC/DC switched-mode power adapter
rated for 100–240 V AC and 50/60 Hz. Every detector is delivered with a set of two power
cords for the following regions:
• EUR (CEE7/7 plug to IEC60320 C13 plug)
• US (NEMA 5-15 plug to IEC60320 C13 plug)
Notice:
For regions with other main plugs/sockets (for example, UK,
Switzerland, Brazil), ensure that you have the appropriate power cords available
at the time of installation and that these power cables are properly grounded and
meet the safety standards that apply in your country. Contact your local
distributor with any questions.
4. The maximum power consumption of the 3465 Detector on full power is < 200 W. The
typical power consumption is < 50 W.
5. Connect the detector to a grounded AC wall socket with a line voltage of 100–240 V AC
(as specified in the sections above) using the supplied power cables. The instrument
should be connected to a protective earth via the socket. Ensure that the detector is placed
in such a way that the main power connection can be reached easily to disconnect it from
the main power by removing the power cable.
December 16, 2021, 715007395 Ver. 00
Page 26