2 Installation
2.1 Storage requirements
The 3465 Detector is shipped to your facility in one box with the following dimensions:
Parameter
Requirement
Height
44.0 cm (17.3 inches)
Width
22.0 cm (8.7 inches)
Depth
43.0 cm (16.9 inches)
Ensure that you have sufficient space to store the packed instrument under the following storage
conditions:
Parameter
Requirement
Storage temperature range
-25 °C to 50 °C (-14 °F to 122 °F)
Storage humidity range
20% to 80%, non-condensing
2.2 Site preparation requirements
Note:
Although not detailed in this document, the installation site must comply with applicable
local laws and regulations with regard to electrical and mechanical installations, building safety,
the use and disposal of potentially hazardous materials and chemicals.
For a successful onsite installation of the instrument, complete the personal computer, laboratory,
electrical and power, and chemicals requirements at your location in advance.
2.2.1 Acquisition computer
The computer used for instrument control requires the following specifications:
• Free LAN port (onboard, PCI, PCI express, or PCI-X)
• Free USB port (required for FW updates)
• Windows 10
For further information about software packages and requirements, visit the Waters website
.
December 16, 2021, 715007395 Ver. 00