XR3D-6 Maintenance & Troubleshooting
5-6
EM-XR3D-6-001-EN.D
©2011 VOTI Inc.
1. Check that the 15-pin connector at the Keypad Control Box is securely in place.
2. Use the DIOX application to verify the Interlock functionality. See
Or check the Keypad for error messages. See
3.3 Control Box Access - Technician.
3. Test for resistance at each interlock switch using a multimeter.
Push the interlock shaft in and check for resistance across the pins. There are 3 pins and
there should be resistance across any 2 of the pins.
There should be a reading greater than zero (0).
4. Gently pull out all the shafts of the Interlock switches to lock the switch closed.
5. Verify if the Interlock Message disappears from the Operator message area.
If the message does not disappear, one of the Interlock Switches may be faulty and must be
replaced.
5.3 Emergency Shut Off
The emergency stop button stops the X-ray source and the motor immediately and
removes all VAC power. The computer, one monitor, and all the 12VDC
(input/output) components remain powered by the UPS until the emergency stop
button is released by turning it clockwise until it pops out.
The UPS produces an audible “beep” to indicate that the main power source has been lost.
EMERGENCY SHUT OFF TROUBLESHOOTING CHART
PROBLEM
CAUSE
CORRECTIVE ACTION
X-Ray source and motor do not stop
when Emergency Stop is depressed.
Cable disconnected at switch or at the
2-Pin connector at the Power
Distribution Unit.
Depress the 2
nd
Emergency Switch to
stop the X-ray source and motor. If this
works, the first switch is faulty or a wire
is disconnected at the switch.
Cable disconnected at Power
Distribution Unit.
Check the cable in the Power
Distribution Unit.
Defective switch.
Check the contact block (the switch).
See
5.3.1 Check the Emergency Switch
Emergency Stop will not extend.
Defective switch.
Replace the switch. See
and Replace the Emergency Stop
Switch
5.3.1 Check the Emergency Switch
Verify the emergency button is fully extended.
1. Push on the button.
2. Turn the button clockwise to loosen and extend the button.
To verify the switch, proceed as follows:
Содержание XR3D-6
Страница 2: ...XR3D 6 Maintenance Troubleshooting ii EM XR3D 6 001 EN D 2011 VOTI Inc...
Страница 5: ...XR3D 6 Maintenance Troubleshooting EM XR3D 6 001 EN D v 2011 VOTI Inc 7 XR3D 6 Safety Features 7 1...
Страница 6: ......
Страница 20: ...XR3D 6 Maintenance Troubleshooting 2 8 EM XR3D 6 001 EN D 2011 VOTI Inc...
Страница 32: ...XR3D 6 Maintenance Troubleshooting 3 12 EM XR3D 6 001 EN D 2011 VOTI Inc...
Страница 48: ...XR3D 6 Maintenance Troubleshooting 3 28 EM XR3D 6 001 EN D 2011 VOTI Inc...
Страница 52: ...XR3D 6 Maintenance Troubleshooting 4 4 EM XR3D 6 001 EN D 2011 VOTI Inc...