XR3D-6 Maintenance & Troubleshooting
EM-XR3D-6-001-EN.D
3-25
©2011 VOTI Inc.
The DIOX window appears.
3. Press the XRAY button (
).
The lights on the top of the scanner will turn red. The DIOX indicators for Red/Green Lights
and X-Ray On/Off will also turn red.
It means that the source is enabled to generate X-Rays but it is not powered. Since the
source is not powered, there is no radiation emitted.
4. Start VOTI's Diagnostic Tool application, which is the software for the X-ray source, located
at
c:\MonoBlock XP\X3825 Spellman*.exe
.
5. Enter parameters if not already displayed.
6. Start XView using
C:\Program Files\DTB\X-view4.2\XView4.2.exe
.
Содержание XR3D-6
Страница 2: ...XR3D 6 Maintenance Troubleshooting ii EM XR3D 6 001 EN D 2011 VOTI Inc...
Страница 5: ...XR3D 6 Maintenance Troubleshooting EM XR3D 6 001 EN D v 2011 VOTI Inc 7 XR3D 6 Safety Features 7 1...
Страница 6: ......
Страница 20: ...XR3D 6 Maintenance Troubleshooting 2 8 EM XR3D 6 001 EN D 2011 VOTI Inc...
Страница 32: ...XR3D 6 Maintenance Troubleshooting 3 12 EM XR3D 6 001 EN D 2011 VOTI Inc...
Страница 48: ...XR3D 6 Maintenance Troubleshooting 3 28 EM XR3D 6 001 EN D 2011 VOTI Inc...
Страница 52: ...XR3D 6 Maintenance Troubleshooting 4 4 EM XR3D 6 001 EN D 2011 VOTI Inc...