XR3D-6 Maintenance & Troubleshooting
3-14
EM-XR3D-6-001-EN.D
©2011 VOTI Inc.
5. In VOTI's Diagnostic Tool application:
a. Click
Program Voltage
to set the anode voltage.
The unit is 0.1 KV. Set the operating voltage at 160 KV, therefore, enter
1600
and click
OK
.
b. Click
Program Current
to set the tube current.
The unit is 1 µA. Set the operating current at 100 µA, therefore, enter
1000
and click
OK
.
6. Select
Tube Re-Seasoning > Profile > Idle *
.
Select the correct idle time depending on the time the unit has not been powered, for
example,
Idle 3-30 Days
.
7. Select
Tube Re-Seasoning > Seasoning > Start Seasoning
.
The
Voltage Set Point
and
Current Set Point
areas flash yellow.