Note:
Also adjustable at Operator Level 3.
OP.DWN
Output Rate
Down
Used to limit the rate at which the output from the PID
terms can decrease (% per second).
Note:
Also adjustable at Operator Level 3.
dIAG
Sublist:
L.BRK.T
Loop Break Time
Used to set the loop break time. If there is a loss of control
within the control loop, this parameter determines the
amount of time before the loop break alarm (
L.BRK
)is
activated.
Note:
Also adjustable at Operator Level 3.
L.BRK.D
Loop Break
Delta PV
Used to set the minimum change in the
PV
that the system
would expect to see within 2x Loop Break Times.
If the controller output is saturated and the
PV
has not
moved by this amount in 2x
L.BRK.T
, then the loop break
alarm will be activated.
Note:
Also adjustable at Operator Level 3.
L.BRK
Loop Break
Detected
Indicates whether a loop break has been detected.
l
Yes
- A loop break has been detected
l
No
- The loop is functioning normally
DEMO
Enable Demo
Mode
Used to turn on a simulation for demonstration purposes.
l
On
- Demo mode is enabled
l
Off
- Demo mode is disabled
Note:
Also adjustable at Operator Level 3.
PROg
List
Note:
The following parameters are only accessible at Configuration Level. When
logged in at Operator Level 3, the parameters displayed are the same as those
available at Operator Level 2. See section 6.3
E.accs
Edit Access
Used to determine the lowest operator level at which edit
permissions are granted. The default value is
Lev2
.
R.ACCS
Run Access
Displays the lowest operator level granted permission to
run a program. This parameter is read only.
RECOV
Recovery
Strategy
Used to detmine how a program behaves if the product is
power cycled.
Ramp
- The controller will ramp to the target setpoint at the
34
Содержание CARBOLITE GERO 3016