Event 2
The number displayed is dependent on the amount of
outputs or alarms configured. This parameter is read only.
P.ADVN
Program
Advance
Enables the operator to force the program to advance to
the next segment.
l
Yes
l
No
Note:
This parameter can only be edited when logged in
at Operator Level 2.
6.3
Create / Edit a Program
Note:
To create a new program, or edit an existing program, the user must first be
logged in at Operator Level
2
(see section 4.0).
To access program setup:
1. From the home display, press the page button twice to reach the
P.Set
option.
2. Press the scroll button to access the list of parameters.
3. Continue to press the scroll button to navigate through the list of parameters.
4. Adjust parameter options using the up and down buttons.
5. When you have finished setting up the program, press the page button multiple
times to return to the top of the menu you are currently in (
S.NUM
>
P.NUM
>
P.SEt
> Home display). Alternatively, the display will automatically return to the home dis-
play after 30 seconds of inactivity.
Note:
The controller will initially scroll through the parameters that apply to the whole
of the program. Continue to press the scroll button to reach parameters specific to a
particular segment within the program.
16