![VERDER CARBOLITE GERO 3016 Скачать руководство пользователя страница 40](http://html.mh-extra.com/html/verder/carbolite-gero-3016/carbolite-gero-3016_installation-operation-and-maintenance-instructions_1003585040.webp)
l
dEG.F
- °Farenheit
l
k
- Kelvin
cs.id
Customer ID
Used to set the product identification number can be
entered if required. This may be used to identify one of
many units for production or quality control systems.
HMI
Sublist:
H.DISP
Home Display
Used to select the information that appears on the home
display.
l
PV.SP
- The home display shows the actual tem-
perature and setpoint.
l
PV.Pt
- The home display shows the actual tem-
perature and program time left.
l
LPU
- The home display shows only the actual tem-
perature.
l
PV1
- The home display shows only the PV for analogue
input 1.
l
PV.PS
- The home display will show the actual tem-
perature and the current running program number and
segment number.
K.LOCK
Keylock
Used to disable the buttons on the front of the controller.
l
On
- The controller buttons are disabled
l
Off
- The controller buttons are enabled
Note:
Once the controller accepts the
On
value, the
buttons will no longer function and the controller must be
linked to iTools for further configuration or use.