Deviation High
- The alarm relay will activate if the
temperature rises above the sum of the Reference and
Deviation values [
REF
±
DEV
], and will remain activated
until the temperature drops below the sum of the Refer-
ence and Deviation values minus any Hysteresis value
set [
REF
+
DEV
-
HYST
]
l
Deviation Low
- The alarm relay will activate if the
temperature drops below the sum of the Reference and
Deviation values [
REF
+
DEV
], and will remain activated
until the temperature rises above the sum of the Refer-
ence and Deviation values plus any Hysteresis value set
[
REF
+
DEV
+
HYST
]
l
Deviation Band
- The alarm relay will activate if the
temperature deviates above or below the Reference
value plus or minus the Deviation value [
REF
±
DEV
], and
will remain activated until the temperature returns
within the deviation band, plus or minus any Hysteresis
value set [
REF
±
DEV
±
HYST
] (e.g. if the target setpoint
is 500°C, the alarm relay will activate if the tem-
perature is outside the range of 450°C and 550°C, and
deactivate when the temperature is between 450°C
and 550°C)
Note:
Also adjustable at Operator Level 3.
Comm
List
s.list
Sublist
Use the up and down arrows to view sublists of additional
parameters. Use the scroll button to select a sublist.
l
MaIN
- Access parameters relating to Main Sublist
l
Nwrk
- Access parameters relating to Network Sublist
mAin
Sublist:
intf
Interface
Displays the type of connection interface.
l
EtH
- Ethernet
l
r485
- RS485
n.sta
Status
Displays the status of communications used by Modbus TCP.
l
OFFL
- Offline and not communicating
l
INIt
- Initialising communications
l
RdY
- Ready to accept connection. Not used by Modbus
TCP.
l
Run
- Ready to accept connections or controller com-
municating
w.time
Watchdog Time
If the communications stop addressing the instrument
longer than this configurable period, the Watchdog Flag will
become active.
37
Содержание CARBOLITE GERO 3016
Страница 51: ...13 0 Navigation Diagrams 13 1 Operator Level 1 13 2 Operator Level 2 51 ...
Страница 52: ...13 3 Operator Level 3 52 ...
Страница 54: ......
Страница 55: ...Notes Service Record Engineer Name Date Record of Work ...