Note:
The minimum possible rate value is 0.1°C per
ramp unit. It is possible to set a ramp rate that far
exceeds the achievable ramp rate of the product. Only
set ramp rates that can be reasonably achieved. Tube
furnaces with ceramic work tubes need slow heating
rates in order to protect the work tube. Please refer to
the "Operation" section of the product manual.
EV.OP
Note:
The 3016 is available with 2 optional relay outputs
which may be configured as event outputs or alarms.
Any parameters above
3
are not applicable to this
controller and should not be used. Do not use values
above the number of events configured. If the 3016 has
been ordered without the relay option, the EV.OP
parameter will not appear.
l
0
- All event outputs are OFF
l
1
- Only event output 1 is ON
l
2
- Only event output 2 in ON
l
3
- Both event output 1 and 2 are ON
tImE
TSP
Used to define the Target Setpoint for that segment i.e.
the temperature (°C) that is to be reached at that point in
the program.
R.TIME
The desired time (in hours, minutes or seconds) that the
user wants the controller to take to reach the desired
setpoint. Dependent on "Ramp Units" parameter.
Maximum time values if the "Ramp Units" are:
l
SECS
= 99:59
l
MINS
= 99:59
l
HRS
= 500.0
EV.OP
Note:
The 3016 is available with 2 optional relay outputs
which may be configured as event outputs or alarms.
Any parameters above
3
are not applicable to this
controller and should not be used. Do not use values
above the number of events configured. If the 3016 has
been ordered without the relay option, the EV.OP
parameter will not appear.
l
0
- All event outputs are OFF
l
1
- Only event output 1 is ON
l
2
- Only event output 2 in ON
l
3
- Both event output 1 and 2 are ON
dwEL
DUR
The length of time (in hours, minutes or seconds) that the
controller should maintain the temperature achieved by the
19