![VERDER CARBOLITE GERO 3016 Скачать руководство пользователя страница 44](http://html.mh-extra.com/html/verder/carbolite-gero-3016/carbolite-gero-3016_installation-operation-and-maintenance-instructions_1003585044.webp)
Graphical Example of Autotuning a Product to 700°C
8. When the autotuning cycle successfully completes, the autotune status (
T.STAT
)
parameter in the
AtuN
list will display
donE
.
Tip:
If you are tuning to optimise the controller for low temperature use, the output
power (
OUT.HI
) value can be reduced before autotune is used. Output power can be
returned to a higher value if the product is subsequently required to operate at higher
temperatures.
The
OUT.HI
parameter can be found within the
OP
sublist in the controller
configuration (see section 9.0).