21. Press the scroll button until
S.TYPE
appears.
22. Press the up and down buttons until
dWEL
appears, then press the scroll button.
23. Press the up and down buttons until the
DUR
value reads
00:30
, then press he scroll
button.
24. Press the scroll button to access values for segment 5.
25. Press the scroll button until
S.TYPE
appears.
26. Press the up and down buttons until
tImE
appears, then press the scroll button.
27. Press the up and down buttons until the
TSP
reads
200°C
then press the scroll but-
ton.
28. Press the up and down buttons until the
R.TIME
value reads
01:20
, then press the
scroll button.
29. Press the scroll button to access values for segment 6.
30. Press the scroll button until
S.TYPE
appears. The default value is
ENd
.
The program is now ready to run.
Segment
1
2
3
4
5
6
Type 'Rate'
Type 'Dwell'
Type 'Step'
Type 'Dwell'
Type 'Time'
Type 'End'
Target 400°C Duration 30 min Target 600°C Duration 30 min Target 200°C End Type 'Reset'