Deviation values [
REF
+
DEV
], and will remain activated
until the temperature rises above the sum of the Refer-
ence and Deviation values plus any Hysteresis value set
[
REF
+
DEV
+
HYST
]
l
Deviation Band
- The alarm relay will activate if the
temperature deviates above or below the Reference
value plus or minus the Deviation value [
REF
±
DEV
], and
will remain activated until the temperature returns
within the deviation band, plus or minus any Hysteresis
value set [
REF
±
DEV
±
HYST
] (e.g. if the target setpoint
is 500°C, the alarm relay will activate if the tem-
perature is outside the range of 450°C and 550°C, and
deactivate when the temperature is between 450°C
and 550°C)
Note:
See section 7.0 for more information on
Hysteresis
.
Note:
If more than one Deviation alarm is specified at
the time of order, multiple
DEV
parameters will be
available. They appear in the order that the alarms were
set up, so if Alarm 1 is an
DeV.H
, and Alarm 2 is an
DeV.L
, the deviation value for the
dEV.H
alarm will be
appear first when scrolling though the parameter list.
LANG
Language
Used to set the language of the scrolling text within the
controller.
l
NONe
- The controller defaults to the last language
selected
l
En
- English
l
Fr
- French
l
dE
- German
l
It
- Italian
l
ES
- Spanish
Note:
Additional controller configuration settings are available in Operator Level
3
and Configuration Level (see section 9.0)
28