![VERDER CARBOLITE GERO 3016 Скачать руководство пользователя страница 17](http://html.mh-extra.com/html/verder/carbolite-gero-3016/carbolite-gero-3016_installation-operation-and-maintenance-instructions_1003585017.webp)
Program Setup
Display
Parameter
Description/ Use
P.NUM
Program
Number
Displays the program number. This parameter is
1
by
default.
P.NAME
Program
Name
Displays the name of the program. This parameter is
P1
by
default.
HB.STY
Holdback
Style
Used to define whether the same holdback settings should
apply for the entire program, or whether they can be
defined for each individual segment.
l
If
PROG
is selected, you will be able to set parameters
for “Holdback Type” and “Holdback Value”.
l
If the
SEGm
style is selected, the
HB.TYP
parameter will
only become available when editing an individual seg-
ment.
HB.TYP
Holdback
Type
Used to define the Holdback Type for the entire program.
l
Off
– holdback does not operate
l
Low
– holdback operates only during heating
l
High
– holdback operates only during cooling
l
Band
– holdback operates during both heating and
cooling
HB.VAL
Holdback
Value
Sets the number of degrees (°C) by which the program can
run ahead of the actual temperature before holdback
operates and prevents further heating or cooling (this
parameter is not visible when the Holdback Type is set as
Off)
RAMP.U
Ramp Units
Used to define whether the ramp units are °C per second
17