Diagnostics and repair
LFXG-H Installation and Operation Guide
91
• Power supply circuit board
Spare parts
Contact VEGA Field Service at
+1 513-272-0131
for parts, service, and
repairs.
Outside the U.S., contact your local VEGA representative for parts, service,
and repairs.
Replacing the CPU or Power supply board
You may have to replace a circuit board if there is damage to one of its
components. Before replacing a circuit board, check the troubleshooting
flowcharts or call VEGA Field Service to be sure a replacement is
necessary.
The sensor EEPROM contains a backup of the CPU board EEPROM. After
replacing the CPU board, you must perform a memory backup to update the
CPU board’s EEPROM with the information in the sensor board EEPROM.
Procedure 5.3: To replace the CPU or power supply board
Note:
Use great care to prevent damage to the electrical
components of the gauge. VEGA recommends appropriate
electrostatic discharge procedures.
1.
Shut off the power to the gauge.
2.
Remove the housing cover.
3.
Remove the plastic electronics cover.
4.
Remove the terminal wiring connector.
5.
Remove the three (3) screws holding the electronics package in
place.
6.
Carefully pull the electronics package out of the housing.
Содержание FiberFlex LFXG-H
Страница 2: ......
Страница 4: ...ii LFXG H Installation and Operation Guide NOTES...
Страница 6: ...iv LFXG H Installation and Operation Guide...
Страница 14: ...Preface xii LFXG H Installation and Operation Guide...
Страница 34: ...Installation 32 LFXG H Installation and Operation Guide FiberFlex mounting...
Страница 91: ...Diagnostics and repair LFXG H Installation and Operation Guide 89 Source Functions...
Страница 100: ...98 LSGH Installation and Operation Guide NOTES...
Страница 101: ......