Calibration
56
LFXG-H Installation and Operation Guide
Procedure 3.7: To calculate the linearizer
5 Calculating the calibration
Procedure 3.8: To calculate the calibration
Repeating the calibration
Typically, the system requires only periodic standardization to compensate
for drifts over time.
However, these events require you to repeat the calibration:
• Measurement of a new process application (contact VEGA for
recommendation)
• Process requires a new measurement span
• Entering a new measurement span setting into the software
• Installing a new radiation source holder
• Moving the gauge to another location
• Changes to the process vessel (example: lining, insulation, or agitator)
• Excessive buildup or erosion of vessel that standardization cannot
compensate for (check standardize gain)
• Standardize gain > 1.2 after a standardization, indicating it made a 20%
adjustment from the last calibration
1.
Select
Calibration | Linearizer Data Pt | Recalculate
.
2.
Click
OK
to proceed with the linearity calculation. The linearizer table
calculates based on the level values.
1.
Select
Calibration | 2 Point Calibration
.
2.
Click
Calculate Results
.
3.
Click
OK
.
4.
Click
OK
.
Содержание FiberFlex LFXG-H
Страница 2: ......
Страница 4: ...ii LFXG H Installation and Operation Guide NOTES...
Страница 6: ...iv LFXG H Installation and Operation Guide...
Страница 14: ...Preface xii LFXG H Installation and Operation Guide...
Страница 34: ...Installation 32 LFXG H Installation and Operation Guide FiberFlex mounting...
Страница 91: ...Diagnostics and repair LFXG H Installation and Operation Guide 89 Source Functions...
Страница 100: ...98 LSGH Installation and Operation Guide NOTES...
Страница 101: ......