Diagnostics and repair
76
LFXG-H Installation and Operation Guide
Some conditions are self-repairing (example: RAM and EEPROM
corruption). Therefore, these may appear in history screens but not
diagnostic screens.
Acknowledging diagnostic alarms
Diagnostic alarms turn off when the problem is solved, except these alarms:
• Source wipe due
• Shutter check due
• Standardize due
Perform the procedure to acknowledge them.
Diagnostic alarm messages
Active alarm messages may appear on the Ohmview 2000 menu if the
alarm condition is selected. You can select individual alarm conditions in the
Alarms | Diagnostic Alarm Enable
tab.
Note:
If the relay is set as a diagnostic alarm, you must
acknowledge all diagnostic alarms to reset the relay.
Содержание FiberFlex LFXG-H
Страница 2: ......
Страница 4: ...ii LFXG H Installation and Operation Guide NOTES...
Страница 6: ...iv LFXG H Installation and Operation Guide...
Страница 14: ...Preface xii LFXG H Installation and Operation Guide...
Страница 34: ...Installation 32 LFXG H Installation and Operation Guide FiberFlex mounting...
Страница 91: ...Diagnostics and repair LFXG H Installation and Operation Guide 89 Source Functions...
Страница 100: ...98 LSGH Installation and Operation Guide NOTES...
Страница 101: ......