Advanced functions
62
LFXG-H Installation and Operation Guide
Gauge Information
Process Variables tab
Table 4.2
Process Variables tab — display values
Value
Description
Min PV
The value, in process units, as entered in the setup tab.
Use this to calculate the measurement span.
Max PV
Counts Low
The temperature and sensor uniformity gain compensated
counts from the sensor at the Cal low level. Determining
the Cal low level occurs during the calibration.
Counts High
The temperature and sensor uniformity gain compensated
counts from the sensor at the Cal high level. Determining
the Cal high level occurs during the calibration.
Sensor Temp Comp
Gain
The current value of the temperature compensation gain.
Use this to adjust for inherent sensor output change with
temperature.
Uniformity Gain
The current of the uniformity gain. Use this to force all
level sensors to output the same counts at a given
radiation field. Most level applications do not use
uniformity gain and use a default of 1.0.