Calibration
LFXG-H Installation and Operation Guide
53
2 Setting the high level and collecting Cal High data
You must:
• Use the gauge to measure the high process condition.
• Enter the actual level.
This sets the gain of the calibration curve. Perform this procedure before or
after setting the low level.
Before starting the Cal High data collection:
Fill vessel with high process, or close the source holder shutter to
simulate high process.
Have the actual level value ready to enter.
Procedure 3.5: To set the cal high level
3 Collecting the linearizer table data
4.
Enter the actual value in engineering units.
5.
Click
OK
.
Note:
You must perform a data collection for the low and high level
within 10 days of each other for a good calibration. The low and
high values must be > 10% of the process span apart for the
most accurate calibration. Increasing the process span usually
increases the gauge accuracy.
1.
Select
Calibration | 2 Point Calibration | Cal High Collect
.
2.
Select
Start
.
3.
Click
Accept
.
4.
Enter the actual level process value in engineering units.
5.
Click
OK
.
Note:
The simple method skips this step.
Содержание FiberFlex LFXG-H
Страница 2: ......
Страница 4: ...ii LFXG H Installation and Operation Guide NOTES...
Страница 6: ...iv LFXG H Installation and Operation Guide...
Страница 14: ...Preface xii LFXG H Installation and Operation Guide...
Страница 34: ...Installation 32 LFXG H Installation and Operation Guide FiberFlex mounting...
Страница 91: ...Diagnostics and repair LFXG H Installation and Operation Guide 89 Source Functions...
Страница 100: ...98 LSGH Installation and Operation Guide NOTES...
Страница 101: ......