Calibration
54
LFXG-H Installation and Operation Guide
This step lets you collect data points between the high and the low
calibration points so the gauge calculates a response curve based on your
data.
Before collecting the linearizer table data:
Prepare to set the level and take data for
10 levels (including the Cal
Low and Cal High levels).
Prepare to enter the levels into the gauge.
You can collect linearizer table data with the data collection for the Cal
Low and Cal High levels.