Calibration
52
LFXG-H Installation and Operation Guide
2.
Enter the actual level.
This sets the low end (sometimes called 0) of the calibration curve. Perform
this procedure before or after setting the high level.
Before starting the Cal Low data collection:
Fill vessel to the low level.
Have the actual level value ready to enter.
Procedure 3.4: To set the cal low level
Note:
Perform a data collection for the low and high level within ten
days of each other for a good calibration. The low and high
values must be > 10% of the process span apart for the most
accurate calibration. Increasing the process span usually
increases the gauge accuracy.
1.
Select
Calibration | 2 Point Calibration | Cal Low Collect
.
2.
Click
Start
.
3.
Click
Accept
.