Calibration
LFXG-H Installation and Operation Guide
43
Procedure 3.1: To calibrate the current loop
1.
Select
Calibration
|
Current Loop Cal
.
2.
Click
Execute.
3.
Click
OK
.
4.
Read the ammeter; enter the actual milliamp reading.
5.
Click
OK
.
6.
Click
YES
if the ammeter reads 4.00 mA or
NO
for any other
reading.
7.
Repeat until the meter reads 4.00 mA. The meter approaches the
4.00 mA successively.
8.
Read the ammeter; enter the actual milliamp reading.
9.
Click
OK
.
10.
Click
OK
.
Note:
If using a voltmeter, calculate the current value.