Diagnostics and repair
LFXG-H Installation and Operation Guide
81
Auxiliary x-ray alarm
To detect x-rays that are causing process changes, a second detector can
be placed outside of the radiation beam of the primary detector. The second
detector only monitors x-ray interference, and has a frequency output that
wires to the auxiliary input of the primary detector.
The primary detector's programming triggers the x-ray alarm when the
counts of the secondary detector are above a threshold.
Call VEGA for more information.
output 10s
before x-ray
current loop output (mA)
Dither level
Dither time
Cycle period
time (ms)
X-ray interference alarm output
Содержание FiberFlex LFXG-H
Страница 2: ......
Страница 4: ...ii LFXG H Installation and Operation Guide NOTES...
Страница 6: ...iv LFXG H Installation and Operation Guide...
Страница 14: ...Preface xii LFXG H Installation and Operation Guide...
Страница 34: ...Installation 32 LFXG H Installation and Operation Guide FiberFlex mounting...
Страница 91: ...Diagnostics and repair LFXG H Installation and Operation Guide 89 Source Functions...
Страница 100: ...98 LSGH Installation and Operation Guide NOTES...
Страница 101: ......