Calibration
Menu:
You get to the calibration menu when Sprint LC’s
LOCKOUT Keyswitch is in the program position. Calibration allows
comparison for Sprint LC to pressure or flow standards. The calibration
menu shows Sprint LC’s actual reading and the pre-programmed target
value the technician compares to the pressure or flow standard. Calibration
should be conducted only by qualified technicians who have proper
training and resources to calibrate Sprint LC.
Page 35, 138-140
Calibration Table:
Technicians don’t need to adjust potentiometers,
dipswitches, or other electrical components to calibrate Sprint LC. All
calibration is done by making adjustments to Sprint LC’s sensor
linearization tables.
Page 142
Concurrent:
Testing multiple ports or multiple parts at the same time is
known as concurrent testing. If the tester has multiple test ports and tests
them all at the same time, it is called a concurrent test.
See Sequential
Contrast:
The contrast of the LCD can be adjusted for the best viewing. It
is adjusted with the contrast knob on the back of Sprint LC.
Page 11, 12
Counters:
Sprint LC records the total number of tests performed (both
pass and fail) and number of rejects (fail only). Running totals are
displayed to operators in the tested and reject boxes on the run display.
The maximum number in either the tested or reject box is 65,535.
Page 17, 29, 82, 85
Counters, how to clear:
Values in the tested and reject boxes on the run
display are reset to zero in the Run Mode Function menu. Press the
ENTER/BACK
keypad, select Clear Counters, and press the
ENTER/BACK
keypad.
Page 29