Page 101
Product Volume
To have Sprint LC calculate the leak rate in cubic centimeters per minute
(ccm), enter the internal volume of the product to be tested plus the fixture
tubing and connections between the tester and the product. During a test,
Sprint LC measures the pressure drop, and calculates the leak rate in units
of cubic centimeters per minute. To turn off the conversion and display
pressure decay set the volume to zero.
Calculation
Sprint LC calculates
the leak rate from the
pressure drop over
time. The leak rate is
calculated by this equation.
Sprint LC automatically adds its own internal volume to the product
volume for the total volume. The tester assumes an atmospheric pressure
of 14.7 psig (1 atmosphere) when calculating the ccm rate.
Turn off flow rate calculation
If the volume is set to zero, then Sprint LC does not convert the leak rate
to cubic centimeters per minute. The test results remain in pressure decay.
Testers with Flow:
Testers with flow sensors
do not have the
Prod Volume
parameter line
because those testers read directly in flow units.
Next Program
The next program is the program that Sprint LC will link to if the product
passes the test as established in the previous program or programs. It is
important to not link a program to itself as an endless test loop results.
If linking to any other program is not wanted, or when at the end of a
linked series of tests, set the next program to zero.
Δ
Pressure * Total Volume
ccm =
Δ
Time
* Atmospheric Pressure
USER
TIP