Flow Calibration
Using a Flow Measuring Meter
In this procedure, use an electronic flow measuring meter as the flow
standard. These standards may have a separate flow transducer that
connects to the main readout and power supply.
Access the flow calibration menu as shown in Figure 79 on page 144.
1.
Set the flow zero point first. Disconnect the air supply and turn the
pressure regulator knob fully out (off). Block the test port to be
sure there is no flow. Press the
DECREMENT
keypad so the
Target
value on the calibration menu reads zero.
2. The
flow
Reading
on the calibration menu should display a value
close to zero. The flow calibration standard connected to the Sprint
LC should also read zero. If not, make sure your setup is correct.
3.
Press and hold the white START switch while momentarily
pressing the
ENTER/BACK
keypad. The flow reading on Sprint LC
should now read .00.
4.
Connect air pressure to the air input connector on the rear of the
tester. Connect the flow measuring meter and the flow restricting
device to the test port. Figure 81.
5.
Slowly adjust the pressure regulator. When the
flow standard
reading matches the target value, press and hold the white START
switch while momentarily pressing the
ENTER/BACK
keypad to
accept this as a calibration number.
6.
Repeat step 5 for each Target value programmed into your Sprint
LC until the maximum range is calibrated.
7.
The last step is to re-check zero. Remove pressure and block the
test port. If the zero reading is not within 25% of tolerance then the
calibration may need to be repeated.
Exit Calibration and allow the tester to sit for 15 minutes before
repeating the calibration. This will allow the valves to cool before
proceeding.