Link Programs, how to:
In the parameters menu, use the
INCREMENT
keypad to select the number of the program to link in the parameter called
Next Program. Set to zero if not linking. See
Parameters Menu
.
Page 39
Linked Programs:
Two or more programs can be linked (consecutively
connected) to perform multiple actions during a single test cycle. For
example, a flow program can be set to follow a pressure decay program so
that when the operator presses the START switch, Sprint LC runs through
a flow test then goes to a pressure decay test. If the product goes out of
parameter at any point in either test, the fail light turns on and the test
ends.
Page 17, 39, 52
LOCKOUT Keyswitch:
This keyed switch locks out operators from
Sprint LC’s setup and calibration functions. The only way to access
programming menu is to use the key and turn the switch to program
(vertical) position.
Page 8-9, 30
Main Digital Readout:
In the center of Sprint LC’s run display the main
digital readout is visible when Sprint LC is in run mode. This readout
shows the pressure and flow values during pressure decay, flow, and burst
testing. The way the main digital readout functions varies with each
testing mode.
Page 13, 62-64
Measurement Units:
See
Units of Measure
.
Menu:
A menu is a list or set of programming options.
See
Setup Menu
.
Modes:
Sprint LC has two modes: run and program.
Occlusion Test:
Occlusion testing is used to test products open passages
by trapping pressure inside a product and then measuring an expected
pressure loss. The abbreviation
Occ
is often used in this guide to refer to
occlusion.
Page 50, 55, 59, 106-111