Decay:
The amount of pressure a product can lose during a test period
before going out of an established tolerance. Also called pressure drop.
See Reject Level
Page 50, 55, 98-99
Digital Readout:
See Main Digital Readout
Dump:
The final phase in a pressure decay test. After Sprint LC
completes a pressure decay test, a valve can be activated to open the
product and Sprint LC’s internal test circuit to atmosphere. The action of
releasing pressure to atmosphere is called dump. If dump is not required
(for instance, if you want to unplug product to vent pressure), set dump to
No.
Dump is valuable if you want to link multiple tests and instruct Sprint LC
to vent pressure between tests.
Page 49, 97
Electronic Regulator:
An optional device that automatically adjusts the
test pressure to a preset value. Multiple test programs with multiple test
pressures may be linked together if this option is installed. It also provides
a way to avoid having the operator manually adjust the pressure for
different test pressures.
Page A1
Event:
The instant pressure drops below the event value during a burst or
creep test. Sprint LC monitors pressure during a burst test or creep test and
holds the maximum pressure value measured before the event (pressure
drop) occurred.
Page 124, A24
Factory-Installed Programs:
Uson ships Sprint LC with four installed
programs. There are a total of 99 test programs available. These programs
can be retained or modified by the customer.
Page 34, 88
Page R5