Page 120
Programming a Flow Test
Follow these steps to program a flow test—
1.
Turn the LOCKOUT keyswitch to program mode.
Test Parameters
is highlighted on the setup menu.
Press the
ENTER/BACK
keypad. Use the
INCREMENT
and
DECREMENT
keypads to select the program
number to edit.
2. Press
NEXT
to highlight
Test
Type
. Use the
INCREMENT
keypad to select
Flow
.
3. Press
the
NEXT
keypad to highlight
Test
Pressure
. Connect a part
with a known acceptable flow rate, then press the
INCREMENT
keypad once. Now use the pressure regulator knob to adjust the test
pressure reading. After the reading is set, push the NEXT keypad.
The highlight bar moves down to Pressure Error.
Read pages 37 - 39 for more information about setting the test
pressure.
4. Use
the
INCREMENT
and
DECREMENT
keypads to set the value for
pressure error. Pressure error is how much overshoot or undershoot
is allowed around the test pressure. Pressure error is needed to
ensure the product reaches the correct pressure before Sprint LC
takes flow readings.
5. Press
the
NEXT
keypad to highlight
Coupling
. Use the GoTo
keypad to enter the pop-up menu and set the values for coupling
time. If no external fixtures are connected to the coupling port, set
the coupling time to zero seconds and the setting to OFF. Press the
BACK
Keypad to exit.