Page 116
Flow Setpoints
Besides pressure error tolerance, flow testing requires two independent
setpoints for flow rates— minimum and maximum flow. To pass a flow
test, the product must be within the pressure tolerance and also within both
flow setpoints at the end of test time. Figures 52 and 53.
If both the pressure error and flow tolerance are exceeded during a test,
Sprint LC shows
P ERR
in the status box because pressure error has
priority over flow tolerance.
FIGURE 56
If the air flow measured by the mass flow
sensor is below the minimum setpoint at the
end of test phase (Figure 56), Sprint LC
turns on the reject light and shows
LOW
in
the status box on the run display.
If dump is active, the dump valve opens and
pressure is vented to atmosphere.
Fill
Test
Zero
Couple
Low Flow Reject
TIME
FLOW
Min Flow
Max Flow
.0 LOW
Test Time Status
PROGRAM 1
ress. Tested Reject
psig 1 1
sccm
Flow is below Minimum setpoint
at the end of the test time.