Page 138
Calibration Access
Procedure
1. Turn the LOCKOUT keyswitch to program position. The startup menu
is displayed.
2. Press the
NEXT
keypad to highlight the calibration option.
3. Press the
ENTER/BACK
keypad to select calibration. A flashing warning
message will appear as shown in Figure 75.
Flashing Access Warning
The warning encountered when reaching the first calibration selection
menu flashes until the operator takes one of two possible actions—
Stop and Leave Calibration
To quit proceeding with calibration, exit by pressing the
ENTER/BACK
keypad when the
Exit Calibration
line is highlighted. Sprint LC defaults to
the
Exit Calibration
line.
Proceed with Calibration
To proceed with calibration, use the
NEXT
keypad to select the
Enter
Calibration
line, then press the
ENTER/BACK
keypad. After the
ENTER/BACK
keypad is pressed, Sprint LC displays the menu in Figure 76.
Verify First
If calibration is within tolerances there is no
need re-set the calibration. The calibration
tolerances can be found on the calibration
report that was shipped with the Sprint LC.