Page 99
Reject Level
The reject level is the maximum pressure drop allowed in a product before
a reject is triggered. The level is the difference between the actual test
pressure (P-1) and the value entered in pressure units. If the pressure is
equal to or greater than the reject level at the end of test phase, the product
passes. The instant the pressure drops exceeds the reject limit in test phase,
the test ends and the reject light turns on. Figure 35. If dump is set to
AUTO
, the dump valve opens and vents the pressure to atmosphere
without waiting for the test phase to finish.
FIGURE 46
Fill
Stabilize
Test
Dump
Zero
Couple
TIME
PRESSURE
Reject Level
PD Test -Pass
Decay stays above
reject level during
test phase
Fill
Stabilize
Test
Dump
Zero
Couple
TIME
PRESSURE
Reject Level
Decay exceeds
reject level during
test phase
PD Test -Fail
P1