Page 49
Stab Time
Lets pressure settle before advancing to test phase in a pressure decay test.
Flow and Burst tests do not have a stabilization time.
Test Time
The period when Sprint LC takes readings to determine the pass/fail
condition of the product under test.
Dump
Dump is the action of venting the product under test to atmosphere after a
test. Dump is used as a way to exhaust pressure between tests and to
ensure there is no safety hazard to an operator in removing a pressurized
product.
Dump can be set to one of three modes—
AUTO
If the
DECREMENT
keypad is held to go below zero, the word
AUTO
appears and dump is in an automatic mode. When dump time is set
to
AUTO
, Sprint LC opens the dump valve and monitors the
pressure trapped in the test circuit. When the pressure declines to
approximately 0.5 psig, the dump valve closes and the dump time
ends.
OFF
Use the
INCREMENT
and
DECREMENT
keypads to set the time to
zero. When time is set to zero, dump is turned off and Sprint LC
will not activate the dump valve.
TIME
Use the
INCREMENT
and
DECREMENT
keypads to select a specific
time up to 999.9 seconds. Sprint LC will dump for the established
amount of time.