Parameters Menu
Items displayed on the Main Parameters menu are summarized below.
Program Number (Top Header)
Program Number
is the number identified (named) set of parameters
currently displayed for one test. When
Program
is highlighted, the
INCREMENT
and
DECREMENT
keypads cycle through the available programs.
Test Type
This parameter establishes the type of test to run. Different testers will
provide various kinds of test types such as pressure decay (PD), flow,
burst, occlusion (OCC), and vacuum (VAC). Individual testers will
incorporate test types based on the specific model of tester purchased.
Test Pressure
Be sure to cap the test port when setting test pressure.
To set the test pressure, first highlight the “Test Pressure” line on the
program menu, then press the
INCREMENT
keypad. Immediately after the
INCREMENT
keypad is pressed, Sprint LC opens the valve that directs
pressure to the test port. Adjust the regulator on the back until the desired
test pressure is displayed on the
Test Press.
Parameter line.
Test Pressure in a Flow Test
For products that have a high flow rate, test pressure can be set with air
flowing through a known good product (at the expected flow rate). This is
called a dynamic test pressure. Whichever way the test pressure is set in a
flow test, remain consistent in how the test pressure is established.