Page 151
Flow Calibration
Using Certified Flow Master
In this procedure, use a fixed restrictor, or flow linearizer, having known
(certified) flow rates at various pressures.
Access the flow calibration menu as shown in Figure 79 on page 144.
1.
The flow zero point is set first, so ensure the air supply to Sprint
LC is disconnected and the pressure regulator knob is adjusted
fully out (off). Block the test port to be sure there is no flow. Now
press the
DECREMENT
keypad so the
Target
value on the
calibration menu reads 0.00.
2.
The flow reading value on the calibration menu should display a
value close to zero. Press and hold the white START switch while
momentarily pressing the
ENTER/BACK
keypad. The flow reading
on the calibration menu should now read 0.00.
3.
Connect air pressure to the air input connector on the rear of the
tester. Connect the flow master device to the test port. Figure 83.
4.
Now, slowly adjust the pressure regulator. Watch the readout on
the
pressure standard
. When the pressure standard is at the right
pressure to generate the target flow through the restrictor, press
and hold the white START switch while momentarily pressing the
ENTER/BACK
keypad to accept this as a calibration number.
5.
Repeat step 4 for each target value programmed into your Sprint
LC until the maximum range is calibrated.
6.
The last step is to re-check zero. Remove pressure and block the
test port. If the zero reading is not within 25% of tolerance then the
calibration may need to be repeated. Exit Calibration and allow the
tester to sit for 15 minutes before repeating the calibration. This
will allow the valves to cool before proceeding.