Plotting Run Display
The data plotting feature requires a large portion of the processor time. In
order to free up needed resources, the run display is modified when
plotting is enabled. Much of the run display is not updated during the test.
¾
The test pressure box displays the set test pressure only. The value
does not change during the test.
¾
The main digital readout (large numbers) stays at 000.
¾
The bar graph is disabled.
The status box is updated and has an additional message,
Plot
.
Plot
is
displayed at the end of the test while the logged data is sent through the
serial port.
In a pressure decay, occlusion, vacuum or burst test the pressure values are
reported. In a flow test both the pressure and flow values are reported.
After all the values are sent to the serial port, the standard data logging
string is also printed. The serial string columns are separated by tabs.
Data Plotting with Concurrent Multi-Channel
In a concurrent Sprint LC data can be collected from only one channel or
test port at a time. Below the
Plot Interval
line in a concurrent Sprint LC is
a second new line. A
Plot Channel
line is added to select the channel to be
plotted. The number set on this line indicates the channel that will be used
to collect the plot values.
Press. Error .5 psig
Coupling GoTo
Plot Interval OFF
Plot Channel 1
Test Params GoTo
Next Program 0
Plot Channel Line
FIGURE 98
Appendix 20