Page 115
Flow Test Stabilization
Stabilize phase is not available in a flow test, so when
setting up a flow test, make sure the test time is long
enough for the product to reach a steady flow rate.
FIGURE 55
If the air pressure at the end of test time is
above the maximum pressure error value,
Sprint LC stops the test and if dump is
active, vents the air to atmosphere (Figure
55). Sprint LC displays
P
ERR
in the
status box on the run display.
USER
TIP
Test
Zero
TIME
PRESSURE
Test Pressure
(- P ERR)
(+ P ERR)
Test Fails with “P-ERR” if the
pressure goes above the
maximum Pressure setpoint.
High Pressure in a Flow Test
.0 P ERR
Test Time Status
PROGRAM 1
Press. Tested Reject
93 psig 1 1
sccm