A-13
A4
TABLE OF COMMANDS
Name
Range
Description and format
90
Normal speed setting
0500 ~ 0970
Normal speed setting [%]
"0800" : 80.0 %
93
Venting valve
operation time setting
0000 ~ 0030
The time to open venting valve since STOP signal is input to
controller.
"0030" : Venting valve will open after 30 seconds passes when
the controller is input STOP signal.
Table A-8
List of RS-485 Settings
Name
Range
Description and Configuration
01
Network ID
0001 / 0032
Sets the Network ID
"0001" : ID = 01 "0032" : ID = 32
02
Multidrop
ON/OFF
0000 / 0001
Sets multidrop ON/OFF setting
"0000" : ON "0001" : OFF
03
Terminator
ON/OFF
0000 / 0001
Terminator ON/OFF setting
"0000" : OFF "0001" : ON