46
SECTION 6
OPERATION
6.6
Software Operation
Setting data can be loaded and overwritten only via RS-485 communication.
Software functions are indicated in the table below (Refer to Table 6-2).
Table 6-2
Software Operating Functions
Function
Descripition
Status
Operation mode
LOCAL, REMOTE or RS-485.
Rotational Speed
Motor current
Run Status
(Note 1)
Operation
START/STOP
Motor ON/OFF
ALARM reset
Reset operation when alarm occur.
Timer
Run time
Maintenance timer
Bearing run time
Setting
Variable Rotation Speed
(Low SPEED)
Changes or sets pump rotational speed.
Relay Outputs
Set actions for Relay outputs.
RS-485
Set communication environment.
Multi drop setting and Network ID setting.
Terminator ON/OFF setting.
History
Alarm History
(Note 1) List of run status.
STATUS
Descripition
NORMAL
Rotating at rated rotation speed
ACCELERATION
Accelerating
IDLE
Coasting
STOP
Stop rotating
E-STOP
Alarm is active (stop rotating)
E-IDLE
Alarm is active (Coasting)
NOTICE
Содержание UTM300B
Страница 8: ...vii Location of label...
Страница 11: ...x This page is intentionally left blank...
Страница 17: ...xvi Table of contents This page is intentionally left blank...
Страница 30: ...CONSTRUCTION AND PRINCIPLE 3 1 Pump Construction 3 2 Principle of Turbo Molecular Pumping 3 3 Control system 3...
Страница 51: ...34 SECTION 5 INSTALLATION This page is intentionally left blank...
Страница 70: ...GAS PURGE 7...
Страница 77: ...60 SECTION 8 TURBO MOLECULAR PUMP RECONDITION This page is intentionally left blank...
Страница 111: ...A 26 Appendix A COMMUNICATIONS This page is intentionally left blank...
Страница 113: ...This page is intentionally left blank...