A-19
A5
COMMAND DESCRIPTION
A5.7 Settings
Commands
SR
Read settings
Reads the set value for a designated settings number.
Sends the 2-character settings number as the sub-command.
Refer to Table A-7 "Table of Settings" for setting number.
SW
Write settings
Overwrites the set value for a designated settings number.Sends the 2-character
settings 4-character set value data as the sub-command.
Refer to Table A-7 "Table of Settings" for setting number and set value.
Answers
SA
Send settings value
Returns the set value for the designated settings number. The 2-character settings
4-character set value is returned as the sub-command.
SV
Invalid setting number
This answer is returned if the designated settings number is invalid.
Returns the 2-character settings number as the sub-command.
Commands
SU
Read user memo
Reads the character string in the user memo.
SX
Write user memo
Overwrites the character string in the user memo.
Sends the 20 characters to set in the user memo as the sub-command.
If less than 20 characters are set, the remaining characters are filled with spaces.
Be sure to always send 20 characters.
Answers SF
Send user memo
Returns as a sub-command the set user memo character string or the 20-character
user memo character string overwritten by the SX command.
Commands SG
Default setting
Restore default setting described in Table A-2 *17.
This setting is reflected after power supply reclosing.
Answer SH
Default setting execute
After power supply reclosing, turbo molecular pump starts up by default setting.
A5.8 Shared Answer
Answer AN
Invalid Command
Answer returned by the power supply after it receives an invalid command.
Содержание UTM300B
Страница 8: ...vii Location of label...
Страница 11: ...x This page is intentionally left blank...
Страница 17: ...xvi Table of contents This page is intentionally left blank...
Страница 30: ...CONSTRUCTION AND PRINCIPLE 3 1 Pump Construction 3 2 Principle of Turbo Molecular Pumping 3 3 Control system 3...
Страница 51: ...34 SECTION 5 INSTALLATION This page is intentionally left blank...
Страница 70: ...GAS PURGE 7...
Страница 77: ...60 SECTION 8 TURBO MOLECULAR PUMP RECONDITION This page is intentionally left blank...
Страница 111: ...A 26 Appendix A COMMUNICATIONS This page is intentionally left blank...
Страница 113: ...This page is intentionally left blank...