Page 25
flow is stopped automatically, process is completed, and it is possible
to open the door and take out the sterilized goods from the chamber.
3.17.4
Exhaust
When the timed sterilization cycle is complete, the unit enters the
exhaust stage, provided that a program other than the liquid program
was selected. The steam is exhausted from the chamber, bringing the
internal pressure down to atmospheric pressure.
Содержание LabSci 11-V
Страница 19: ...Page 18 8 Right service door lock...
Страница 22: ...Page 21 10 Drain outlet...
Страница 34: ...Page 33 Attention This device is not a medical device and not intended for medical use...
Страница 35: ...Page 34 During the process the stages of the cycle will be displayed on the screen...
Страница 74: ...Page 73 1 Open the printer s cover door 1 by pulling it up see Fig 2 Fig 1...
Страница 79: ...Page 78...
Страница 90: ...Page 89 Attention Use protective gloves in order not to burn your hands with the hot steam Safety valve ring...